U.S. Department of Transportation
Federal Highway Administration
1200 New Jersey Avenue, SE
Washington, DC 20590
202-366-4000


Skip to content
Facebook iconYouTube iconTwitter iconFlickr iconLinkedInInstagram

Federal Highway Administration Research and Technology
Coordinating, Developing, and Delivering Highway Transportation Innovations

Report
This report is an archived publication and may contain dated technical, contact, and link information
Publication Number: FHWA-HRT-04-124
Date: April 2005

Lab & Field Testing of AUT Systems for Steel Highway Bridges

6. RESULTS (cont'd)

The results of the fourth set of field tests at HSS are summarized in table 11. There were 10 field specimens inspected; 5 of the 10 specimens were designated as FCMs. Table 11 indicates that three FCM and three non-FCM specimens were accepted with no rejectable defects by all of the inspection methods. The 76.2-mm- (3-inch-) thick field specimen FG40M-TF1-Curved-FCM (figures 120 and 121) is a plate specially fabricated for a railroad bridge. Unlike conventional plates that are generally fabricated flat, specimen FG40M-TF1-Curved-FCM was fabricated at a right angle. RT accepted the weld with no rejectable defects. Since specimen FG40M-TF1-Curved-FCM is an FCM, it is required to be inspected by UT. Two rejectable indications were detected by manual UT and one rejectable indication was detected by AUT. Figure 122 shows a radiographic image of the weld section where UT detected rejectable indications. Figures 123 through 130 show the P-scan images created by scanning the weld with a 45-degree probe. The rejectable indication is located between Y = 228.6 mm and Y = 457.2 mm (Y = 9 inches and Y = 18 inches) (figures 125 and 126). According to table 6.2 in the AASHTO/AWS D1.5M/D1.5: 2002 Bridge Welding Code,(1) a 76.2-mm- (3-inch-) thick plate is required to be scanned by 45-degree and 70-degree probes. However, a P-scan inspection of the weld with a 70-degree probe was not performed because of scheduling constraints. The 50.8-mm- (2-inch-) thick field specimen FG1A-TF2-BottF-FCM (figures 131 and 132) was rejected by all three inspection methods. RT rejected the specimen based on a rejectable slag inclusion in section A‑B. According to table 6.2 in the AASHTO/AWS D1.5M/D1.5: 2002 Bridge Welding Code,(1) a 50.8-mm- (2-inch-) thick plate must be scanned by 60-degree and 70-degree probes. Manual UT and AUT rejected the specimen based on an indication detected with the 60-degree and 70-degree probes. The P-scan images of specimen FG1A-TF2-BottF-FCM, scanned with 60‑degree and 70-degree probes, are shown in figures 133 through 135. Specimen G3VHW-CF1-BottF (Figure 136) was rejected by RT based on a slag inclusion found at location Y = 250.952 mm (Y = 9.88 inches). The radiographic image of this weld section is shown in figure 137, but the defect is not clearly visible. Manual UT and AUT inspection with a 70-degree probe found an acceptable indication at Y = 250.952 mm (Y = 9.88 inches). Figure 138 shows the P-scan images from Y = 0 mm to Y = 304.8 mm (Y = 0 inches to Y = 12 inches) for specimen G3VHW-CF1-BottF. Field specimen G5VHW-CF1-BottF (figures 139 and 140) was rejected by RT, but was accepted by manual UT and AUT. RT found a 6.35-mm- (0.25-inch-) long slag inclusion at Y = 639.83 mm (Y = 25.29 inches) (Figure 141). The defect is not clearly visible in the radiographic image because of the poor quality of the radiographic film. Specimen G5VHW-CF1-BottF was not required to be tested by manual UT; however, the fabricator used UT to determine the depth of the defect for repair purposes. Manual UT and AUT identified the defect found by RT and this indication was found to be acceptable. The P-scan images from segment 4 in specimen G5VHW-CF1-BottF were produced using 45-degree and 70-degree probes (figures 142 through 144). Figure 142 illustrates that the indication detected by AUT was within code requirements.

The results of the field tests at Stupp are summarized in table 12. There were four field specimens tested, with all three inspection methods accepting all four specimens.

The photo shows a plan view of field specimen FG40M-TF1-Curved-FCM with the weld oriented vertically. The specimen is comprised of curved plates of unequal thickness and equal width. The thicker plate is to the right in the photo. A man is crouched in front of the specimen. He has the P-scan system mounted on the specimen and is managing the ultrasonic transcducer with both hands. He is looking at the lap top computer positioned on a box in front of him. A large plastic box is open beside him with cords coming out of it to power strips and the computer. He is wearing jeans, a plaid shirt and a white hardhat.
Figure 120. Field specimen FG40M-TF1-Curved-FCM: Top view of joint.

The photo shows a side view of field specimen FG40M-TF1-Curved-FCM with the weld oriented vertically. The specimen is comprised of curved plates of unequal thickness and equal width. The thicker plate is to the right in the photo.
Figure 121. Field specimen FG40M-TF1-Curved-FCM: Side view of joint.

The radiographic image between markers B and C does not show any discontinuities.
Figure 122. Radiographic image of field specimen FG40M-TF1-Curved-FCM: Section B-C.

Figure 123 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 0 and 228.6 millimeters (0 and 9 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 123. P-scan images of field specimen FG40M-TF1-Curved-FCM: From TSC side of centerline between 0 and 228.6 mm (0 and 9 inches).

Figure 124 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 0 and 228.6 millimeters (0 and 9 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 124. P-scan images of field specimen FG40M-TF1-Curved-FCM: From BSC side of centerline between 0 and 228.6 mm (0 and 9 inches).

Figure 125 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one rejectable indication between 228.6 and 457.2 millimeters (9 and 18 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 125. P-scan images of field specimen FG40M-TF1-Curved-FCM: From TSC side of centerline between 228.6 and 457.2 mm (9 and 18 inches).

Figure 126 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one rejectable indication between 228.6 and 457.2 millimeters (9 and 18 inches) from the datum. Note that 1 inch equals 25.4 millimeters.Figure 126. P-scan images of field specimen FG40M-TF1-Curved-FCM: From BSC side of centerline between 228.6 and 457.2 mm (9 and 18 inches).

Figure 127 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 457.2 and 685.8 millimeters (18 and 27 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 127. P-scan images of field specimen FG40M-TF1-Curved-FCM: From TSC side of centerline between 457.2 and 685.8 mm (18 and 27 inches).

Figure 128 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 457.2 and 685.8 millimeters (18 and 27 inches) from the datum. Note that 1 inch equals 25.4 millimeters.Figure 128. P-scan images of field specimen FG40M-TF1-Curved-FCM: From BSC side of centerline between 457.2 and 685.8 mm (18 and 27 inches).

Figure 129 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 685.8 and 993.8 millimeters (27 and 39.125 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 129. P-scan images of field specimen FG40M-TF1-Curved-FCM: From TSC side of centerline between 685.8 and 993.8 mm (27 and 39.125 inches).

Figure 130 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 685.8 and 993.8 millimeters (27 and 39.125 inches) from the datum. Note that 1 inch equals 25.4 millimeters.

Figure 130. P-scan images of field specimen FG40M-TF1-Curved-FCM: From BSC side of centerline between 685.8 and 993.8 mm (27 and 39.125 inches).

The photo shows a plan view of field specimen FG1A-TF2-BottF-FCM with the weld oriented horizontally. The specimen is comprised of plates of equal thickness and width.
Figure 131. Field specimen FG1A-TF2-BottF-FCM: Top view of joint.

The photo shows a side view of field specimen FG1A-TF2-BottF-FCM with the weld oriented vertically. The specimen is comprised of plates of equal thickness and width.
Figure 132. Field specimen FG1A-TF2-BottF-FCM: Side view of joint.

Figure 133 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one rejectable indication between 0 and 254 millimeters (0 and 10) inches from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 133. P-scan images of field specimen FG1A-TF2-BottF-FCM: From TSC side of centerline using 60-degree probe.

Figure 134 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one rejectable indication between 0 and 254 millimeters (0 and 10 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 134. P-scan images of field specimen FG1A-TF2-BottF-FCM: From TSC side of centerline using 70-degree probe.

Figure 135 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 0 and 254 millimeters (0 and 10 inches) from the datum. Note that 1 inch equals 25.4 millimeters.

Figure 135. P-scan images of field specimen FG1A-TF2-BottF-FCM: From BSC side of centerline using 70-degree probe.

The photo shows a plan view of field specimen G3VHW-CF1-BottF-FCM with the weld oriented horizontally.
Figure 136. Field specimen G3VHW-CF1-BottF: Top view of joint.


The radiographic image does not show any discontinuities between markers A and B.
Figure 137. Field specimen G3VHW-CF1-BottF: Radiographic image of section A-B.

Figure 138 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one acceptable indication between 0 and 304.8 millimeters (0 and 12 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 138. Field specimen G3VHW-CF1-BottF: P-scan images from TSC side of centerline using 70-degree probe.

The photo shows a plan view of field specimen G5VHW-CF1-BottF with the weld oriented horizontally. The specimen is comprised of plates of unequal thickness. The thicker plate is at the top of the photo, and the thinner plate is at the bottom of the photo.
Figure 139. Field specimen G5VHW-CF1-BottF: Top view of joint.

The photo shows a side view of field specimen G5VHW-CF1-BottF with the weld oriented vertically. The specimen is comprised of plates of unequal thickness. The thicker plate is on the left side of the photo, and the thinner plate is on the right side of the photo.
Figure 140. Field specimen G5VHW-CF1-BottF: Side view of joint.

The radiographic image does not show any discontinuities between markers B and C.
Figure 141. Field specimen G5VHW-CF1-BottF: Radiographic image of section B-C.

Figure 142 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display identifies one acceptable indication between 482.6 and 666.75 millimeters (19 and 26.25 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 142. P-scan images of field specimen G5VHW-CF1-BottF: From TSC side of centerline using 45-degree probe.

Figure 143 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 482.6 and 666.75 millimeters (19 and 26.25 inches) from the datum. Note that 1 inch equals 25.4 millimeters.

Figure 143. P-scan images of field specimen G5VHW-CF1-BottF: From BSC side of centerline using 45-degree probe.


Figure 144 shows a color-coded image created by the P-scan system which includes the C-scan, B-scan, side view, and response amplitude profile of the weld. The vertical and horizontal axes of the C-scan, B-scan, and side views represent weld dimensions in inches. The vertical and horizontal axes of the amplitude response graph are response amplitude in decibels and distance in inches, respectively. The P-scan image also contains a bar graph relating response magnitude to a series of colors. The colors range from red, which indicates a high amplitude response, to purple, which indicates a low amplitude response. The display does not identify any indications between 482.6 and 666.75 millimeters (19 and 26.25 inches) from the datum. Note that 1 inch equals 25.4 millimeters.
Figure 144. P-scan images of field specimen G5VHW-CF1-BottF: From BSC side of centerline using 70-degree probe.

For comparison, the laboratory and field specimens with rejectable defects were grouped according to plate thickness in 25.4-mm (1-inch) increments (tables 13 and 14). These groupings summarized the results and helped determine the relationship between the effectiveness of each inspection method with respect to the plate thickness. The first column of each table shows the thickness range of the grouped plates. The second column shows the total number of specimens inspected. The third column indicates the number of specimens accepted with no rejectable defects by all three inspection methods. The fourth column shows the number of specimens rejected with rejectable defects by at least one inspection method. The fifth column identifies the rejected specimen. The sixth column shows the number of rejectable defects found. The seventh, eighth, and ninth columns indicate the inspection methods that rejected the particular specimen. The checkmark (check mark) stands for "Rejected," indicating that the specimen was rejected by the employed inspection method; X stands for "Accepted," indicating that the specimen was accepted by the employed inspection method; and DWC stands for "Detected/Within Code," indicating that the employed inspection method identified an indication that was within code requirements. Table 13 shows that the three inspection methods produced consistent results in the laboratory. In table 14, there are 8 field specimens with thicknesses ranging from 0 to 50.8 mm (0 to 2 inches) that contained 12 rejectable defects. RT rejected all eight specimens, while UT rejected only three specimens. Note that UT identified defects in the five remaining specimens, but found that the defects were acceptable. In addition, 5 specimens with thicknesses ranging from greater than 50.8 to 101.6 mm (2 to 4 inches) containing 8 rejectable defects also were identified. RT rejected only two specimens, while UT rejected all five specimens. Note that FG36K-TF2-TopF-FCM was not rejected by AUT since the defect was located at the curvature region of the width transition and the P-scan system was not configured to inspect this type of region.

Tables 15 and 16 compare the results of the RT and AUT inspections. The tables are arranged to illustrate: (1) the total number of specimens inspected within each 25.4-mm (1-inch) thickness increment, (2) the total number of specimens that were rejected by at least one inspection technique, (3) the number of specimens rejected by each inspection method, and (4) the number of specimens accepted with identifiable indications.

Table 13. Consolidating the results of laboratory testing using laboratory specimens with rejectable defects.

Thickness Range (inch)

Total No. of Laboratory Specimens

No. of Laboratory Specimens Accepted

No. of Laboratory Specimens Rejected by at Least One Technique

Rejected Laboratory Specimen

Ind. No.

Rejected by RT

Rejected by UT

Rejected by AUT

0-1

6

0

6

S033 (0.5" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

S034 (0.5" thick)

1

check mark

check mark

check mark

2

check mark

check mark

X1

S125 (1" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

3

check mark

check mark

check mark

S126 (1" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

3

check mark

check mark

check mark

4

check mark

DWC

check mark

S135 (1" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

3

check mark

check mark

check mark

S136 (1" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

>1-2

4

2

2

S132 (1.5" thick)

1

check mark

check mark

check mark

S133 (1.5" thick)

1

check mark

check mark

check mark

1AUT is not configured to detect transverse crack.
DWC: Detected/Within Code

1 inch = 25.4 mm

Table 14. Consolidating the results of field testing using field specimens with rejectable defects.

Thickness Range (inch)

Total No. of Field Specimens

No. of Field Specimens Accepted

No. of Field Specimens Rejected by at Least One Technique

Rejected Field Specimen

Ind. No.

Rejected by RT

Rejected by UT

Rejected by AUT

0-1

5

3

2

G5G-TF1-TopF (1" thick)

1

check mark

DWC

DWC

2

check mark

DWC

DWC

TP2 (1" thick)

1

check mark

check mark

check mark

2

check mark

check mark

check mark

3

check mark

check mark

check mark

4

check mark

check mark

check mark

>1-2

23

17

6

FG26G-TF2-BottF-FCM (1.75" thick)

1

check mark

DWC

DWC

FG16D-TF1-BottF-FCM (2" thick)

1

check mark

DWC

DWC

AWS-FCM-02-6A (1.5" thick)

1

check mark

check mark

check mark

FG1A-TF2-BottF-FCM (2" thick)

1

check mark

check mark

check mark

G3VHW-CF1-BottF (1.75" thick)

1

check mark

DWC

DWC

G5VHW-CF1-BottF (1.75" thick)

1

check mark

DWC

DWC

>2-3

13

9

4

FG36K-TF2-TopF-FCM (3" thick)

1

check mark

check mark

X*

2

X

check mark

check mark

FG37K-TF3-BottF-FCM (3" thick)

1

X

check mark

check mark

FG38K-TF2-TopF-FCM (3" thick)

1

check mark

check mark

check mark

2

X

check mark

check mark

FG40M-TF1-Curved-FCM (3" thick)

1

X

check mark

check mark

2

X

check mark

check mark

3-4

3

2

1

G2G-CF1-BottF-FCM (3.15" thick)

1

X

check mark

check mark

*AUT is not configured to detect transverse defects or inspect width transition regions.

1 inch = 25.4 mm

Table 15. Assessing the detectability and rejectability of three inspection methods in the laboratory.

Thickness Range (inch)

No. of Laboratory Specimens With Rejectable Defects

No. of Specimens Rejected by AUT

No. of Specimens Accepted by AUT Within Code

0-1

6

6 (100%)

0

1-2

2

2 (100%)

0

Table 16. Assessing the detectability and rejectability of three inspection methods in the field.

Thickness Range (inch)

Total No. of Field Specimens

No. of Field Specimens With Rejectable Defects

No. of Specimens Rejected by RT

No. of Specimens Rejected by AUT

No. of Specimens Accepted by RT Within Code

No. of Specimens Accepted by AUT Within Code

0-1

5

2

2 (100%)

1 (50%)

0

1

>1-2

23

6

6 (100%)

2 (33%)

0

4

2-3

13

4

2 (50%)

4 (100%)

0

0

3-4

3

1

0 (0%)

1 (100%)

0

0

ARTICULATION ANGLES

To supplement field testing and establish the importance of the code-prescribed transducer articulation, specimen S033 was tested at a series of fixed angles (0 degrees, 2.5 degrees, 5 degrees, 7.5 degrees, and 10 degrees). The angles were measured with respect to a line normal to the weld centerline. The main purpose of articulating the transducer is to obtain the highest possible amplitude echo from a given defect. To accomplish this, the transducer must be articulated so that the direction of the shear wave propagating within the specimen is perpendicular to a plane of the defect. As mentioned earlier, specimen S033 (shown in figures 7 through 10) contained manufactured crack-like defects oriented along the centerline in the weld bevels.

The testing is performed by the P-scan system using a test frame (Figure 145) and a series of wedges designed to keep the transducer at a fixed angle during scanning (Figure 146). Figure 146 shows an angled wedge attached to a vertical sliding bar which is oriented transverse to the weld axis. The transducer is placed against the angled wedge, holding the articulation angle fixed relative to the weld axis. As the sliding bar is advanced along the length of the weld, the transducer is held at the fixed angle of interest.

The maximum echo amplitudes from the defect were determined from the P-scan images and were plotted against the transducer articulation angles (Figure 147). The abscissa in Figure 147 represents the transducer articulation angles, and the ordinate represents the indication rating (d). The indication rating of negative dB is indicative of a high-amplitude echo and the indication rating of positive dB is indicative of a low-amplitude echo. It is evident from Figure 147 that the amplitude of the echo decreases from a high value of -4 dB to a low value of +14 dB as the articulation angle increases from 0 to 10 degrees, respectively. The 0-degree articulation angle, which produces the highest indication rating of -4 dB, reveals that the direction of the incident traveling wave is aligned normal to the plane of the crack thus reflecting the highest amount of ultrasonic energy. For articulation angles other than 0 degrees, the incident traveling wave is no longer perpendicular to the plane of the crack, leading to less ultrasonic energy reflection. The difference in the echo amplitude between the articulation angles of 0 degrees and 2.5 degrees, 2.5 degrees and 5 degrees, 5 degrees and 7.5 degrees, 7.5 degrees and 10 degrees are 5 dB, 6 dB, 5 dB, and 2 dB, respectively. The 2-dB to 6‑dB difference in the echo amplitude is significant in rejecting or accepting a weld based on the indication rating criteria set forth in the UT acceptance-rejection criteria in tables 6.3 and 6.4 in the AASHTO/AWS D1.5M/D1.5: 2002 Bridge Welding Code.(1) Therefore, a rejectable defect may be misrepresented as an acceptable indication if articulation of the transducer is eliminated.

The photo shows the transducer articulation test setup. In this setup, the magnetic base is positioned in the "Both" configuration while the transducer is held at a constant 10-degree angle relative to the transverse axis of the weld by an angular wedge. The angular wedge is attached to an aluminum frame, which facilitates movement of the transducer along the weld length.
Figure 145. Transducer articulation testing: Test setup.

 

The photo shows a plan view of five angular wedges (0, 2.5, 5, 7.5, and 10 degrees). The angle of the wedge increases from the right side to the left side of the photo. Therefore, the 0-degree wedge is on the right, and the 10-degree wedge is on the left.
Figure 146. Transducer articulation testing: Various articulation angle wedges.

 

This graph shows the relationship of the transducer articulation angle to the amplitude of a reflected signal. The horizontal axis is transducer articulation angle in degrees ranging from 0 to 12. The vertical axis is the indication rating in decibels ranging from negative 8 to positive 16. The negative decibel rating indicates a high amplitude echo, while the positive rating is a low amplitude echo. The graph indicates that the 0-degree and 10-degree transducers produce the largest and smallest amplitude responses, respectively.
Figure 147. Influence of transducer articulation angle on the maximum amplitude of the reflected signal.
Previous    Table of Contents    Next

 

Federal Highway Administration | 1200 New Jersey Avenue, SE | Washington, DC 20590 | 202-366-4000
Turner-Fairbank Highway Research Center | 6300 Georgetown Pike | McLean, VA | 22101