Skip to contentUnited States Department of Transportation - Federal Highway AdministrationSearch FHWAFeedback

Pavements

<< PreviousContentsNext >>

Use of Magnetic Tomography Technology to Evaluate Dowel Placement

FIELD TESTING AND DEMONSTRATIONS

Several site visits were made to demonstrate MIT Scan-2 to State DOT and FHWA personnel, as well as to contractors and representatives of concrete paving trade organizations. These site visits were also instrumental in fulfilling various aspects of field evaluation of MIT Scan-2, including the following:

  • Compare MIT Scan-2 results with those provided by other NDT devices, including GPR and cover meters.
  • Evaluate functional aspects of MIT Scan-2, such as battery life, data storage capacity, and the range of operating environmental conditions.
  • Identify any need for software modifications or enhancements to improve how the operation of MIT Scan-2 fits U.S. practices.
  • Determine productivity of MIT Scan-2.
  • Determine reliability of data storage.

South Carolina was the first State to use MIT Scan-2 in an actual construction project to monitor dowel alignment. Reconstruction of I-95 through Florence was the first project in South Carolina that was constructed using a DBI. The dowel alignment was closely monitored using MIT Scan-2. That experience provided valuable preliminary information on the operation of MIT Scan-2 and the monitoring of dowel alignment. Based on the information obtained through field testing and demonstrations, experience from other projects involving MIT Scan-2, and the lessons learned from the South Carolina experience, guidelines for evaluating dowel alignment using MIT Scan-2 were developed and are provided in Appendix B, Guidelines for Evaluating Dowel Alignment Using the MIT Scan-2 Device.

Site Visits

The States visited include Iowa, Kansas, Minnesota, Missouri, Nevada, North Carolina, and Pennsylvania. The sites were selected based mainly on the agency interest and geographical distribution. In many cases, the site visits were used to provide answers to the questions that had been raised about dowel bar alignment or tie bar position.

Iowa

In Iowa, the DOT had some concern over the dowel bar alignment when cracking occurred along the ends of the dowels at a few transverse joints in a newly constructed JPCP. The MIT Scan-2 results showed that the dowel bar alignments were in fact very good. Testing the four joints in question and two longitudinal joints (for tie bars) took less than 15 min, and the results conclusively showed that the dowel bars are in very good alignment. No other method of testing would have been able to provide the same information so easily in such a short time.

Kansas

In Kansas, a project under construction was suspected of having problems with tie bar position. The DOT conducted GPR testing to determine the extent and severity of the problem, and this provided the opportunity to compare MIT Scan-2 and GPR results. A short section of the project was scanned using MIT Scan-2 to compare the results with those obtained using GPR. The very first joint scanned showed a significant discrepancy between GPR and MIT Scan-2 results. An open-air demonstration was conducted to verify that the MIT Scan-2 results are accurate. A scan of several joints in a row revealed that for some of the joints, the results compared very well.

Table 8 provides a summary of the comparison for 10 joints. The joints were matched by examining the bar location pattern. The variations in the tie bar position served as a fingerprint to match the joints between GPR and MIT Scan-2 testing results. Since consecutive joints were tested, only the location of the first joint tested using MIT Scan-2 needed to be located in the GPR test results. Table 8 shows that where good agreement exists in the detected bar positions, good agreement in bar depths is observed.

Table 8. Comparison of GPR and MIT Scan-2 Results From the Kansas Project
SlabApproximate joint location (station), mGPRMIT Scan-2
Bar loc. cmDist. to next bar, cmCover depth, mmBar loc. cmCover depth, mm
133,330.024.2
105.2
176.0
233.8
291.1
383.1
465.8
80.9
70.8
57.7
57.3
91.9
82.7
34.8
143
141
145
147
154
152
149

105.2
177.0
234.9
293.3

140
148
151
158
233,335.0572.1
651.8
732.8
813.4
883.0
79.6
81.0
80.5
69.6
115.2
150
145
145
141
137
572.1
651.9
734.3
814.9
884.8
152
144
142
140
137
333,340.01,024.7
1,082.0
1,142.2
1,220.7
1,307.6
1,397.6
1,495.5
57.2
60.2
78.4
86.9
89.9
97.9
23.3
137
136
135
131
140
140
158

1,082.0
1,142.4
1,220.2
1,308.7
1,398.1

139
136
139
141
146
433,345.21,562.8
1,653.2
1,722.7
1,790.6
1,872.9
1,967.0
90.3
69.5
67.8
82.3
94.0
34.0
140
141
143
141
139
155
1,562.8
1,654.0
1,724.6
1,792.3
1,875.3
145
149
149
147
141
533,350.22,115.7
2,181.1
2,256.5
2,324.0
2,397.8
2,454.7
2,478.5
65.3
75.4
67.4
73.8
56.8
23.8
102
126
125
125
126
125
154
2,115.7
2,213.4
2,307.1
2,398.7
2,481.5
144
149
142
144
138
633,355.22,569.6
2,626.5
2,697.3
2,771.2
2,820.8
2,881.9
56.8
70.8
73.8
49.6
61.0
128
129
128
129
130
129
2,569.6
2,667.7
2,761.2
2,853.1
2,937.4
150
151
147
148
144
733,360.23,094.9
3,168.4
3,352.6
73.4
184.2
119
124
105
3,094.9
3,174.2
3,267.4
3,344.3
3,424.7
132
139
133
140
136
833,365.13,540.7
3,614.1
3,685.4
3,787.6
3,916.4
73.3
71.3
102.1
128.8
102
119
119
119
119
3,540.7
3,622.6
3,706.1
3,790.3
136
135
132
134
933,370.14,018.8
4,086.3
4,163.0
4,215.3
4,215.8
4,288.4
4,366.4
4,438.1
67.4
76.7
52.2
0.5
72.5
78.0
71.6
102
128
130
129
129
128
128
129

4,086.3
4,216.7
4,301.2
4,424.1

131
134
136
121
1033,375.14,570.8
4,652.7
4,724.7
4,843.5
4,914.3
81.8
72.0
118.7
70.8
126
97
105
127
128
4,570.8
4,701.1
4,785.8
4,932.9
131
134
136
121

Because MIT Scan-2 results are a direct interpretation of the magnetic signals detected during testing, the bar location reported by the device is highly reliable. On the other hand, the GPR results depend on material properties used in the data evaluation. Since the material properties can vary significantly along the length of a project, localized errors are possible. Some of the bar positions reported by GPR appear to be in error. For example, in slab 7, GPR reported only three tie bars, and the bar spacing in slabs 8, 9, and 10 is highly variable, ranging from 0.5 to 129 cm (0.2 to 51 in.). Results are shown graphically in Figure 36.

Figure 36. Comparison of GPR and MIT Scan-2 results from Kansas.
Comparison of GPR and MIT Scan-2 results from Kansas. Scatterplot graph of data contained in Table 8, the bar locations and their cover depths as determined using GPR and MIT Scan-2.
Minnesota

Prior to laboratory testing at the MnRoad facility, a short presentation was made to the MnRoad and MnDOT staff, followed by a demonstration. MnDOT has a cover meter, and the results of MIT Scan-2 were compared with the data collected using the cover meter. The results were described in the Literature Review section. The depth measurements were nearly identical. To measure dowel alignment using a cover meter, the ends of the dowel bar needed to be located and marked manually by finding the location where the signal drops off abruptly. While accurate results could be obtained using a cover meter, testing a large number of bars using this device is not practical. As noted in the Literature Review, cover meters are subject to the same limitations as MIT Scan-2: presence of other metal objects will affect the measurement results.

Missouri

The Missouri DOT was interested in a demonstration of MIT Scan-2, and a site visit could be conveniently coordinated with the visits to Kansas, Iowa, and Minnesota. A short presentation was made at the DOT office, followed by an open-air demonstration and a field demonstration. The field demonstration was made on an on-ramp under construction near the DOT office. Several transverse joints were scanned, along with one longitudinal joint. The dowel alignment was not a concern on this project. The scan results showed no major problem with dowel alignment. One noticeable feature on this project was that all joints had at least a few uncut ties, clearly visible on the signal intensity plot (Figure 37).

Figure 37. Signal intensity plot showing uncut ties between location 110 and 150 (cm).
Signal intensity plot showing uncut ties between location 110 and 150 (cm).
Nevada

Field tests were conducted on an I-80 reconstruction project in Reno. This was an opportunity to evaluate the feasibility of using MIT Scan-2 to evaluate alignment of dowel bars placed in baskets. This series of tests demonstrated that MIT Scan-2 is a valuable tool for identifying alignment problems of dowel baskets even without the basket software. The test results showed that the problems that develop in dowel baskets tend to be more obvious, larger scale problems that are easy to detect from the graphical output of MagnoProof (Figure 35). In general, the problems appear to be results of the baskets bursting open or deforming during concrete placement due to inadequate anchoring.

On this project, the baskets were originally anchored using only four pins per basket, one at each corner. MIT Scan-2 test results showed that these were not sufficient to hold the basket in place. Many baskets burst open, causing large misalignments. The most common problems observed from this project are shown in Figure 35b. Such problems are also clearly evident in the numerical results obtained using the standard software. The numerical results obtained without the proper calibration must be considered qualitative, but the results still provide an adequate degree of accuracy to clearly delineate the baskets that have problems and those that exhibit very good alignment.

After discovering the problem with the basket anchoring practice, the process was modified to pin every other dowel on each side of the basket (10 pins per basket). This change resulted in immediate and drastic improvement in dowel alignment. A comparison of the dowel alignment trends before and after the change is shown in Figure 38. Severely misaligned dowel bars can cause a joint to lock. Figure 38 shows that anchoring the baskets using only 4 pins per basket caused 11.5 percent of dowel bars to be misaligned by 20 mm (0.78 in.) or more. This is a severe case of dowel misalignment. If uniformly distributed, the 11.5 percent would place at least one severely misaligned dowel bar in every joint. By contrast, the baskets anchored using 10 pins per basket had exceptional alignment. No bars were misaligned by more than 20 mm (0.78 in.), and only 0.2 percent of the bars were misaligned by more than 15 mm (0.59 in.). These results represent about the best dowel alignment that could possibly be achieved in the field using either baskets or DBI. All sections anchored using 10 pins per basket showed similar results on this project.

Figure 38. Comparison of dowel alignment for baskets anchored using 4 pins/basket and 10 pins/basket.
Comparison of dowel alignment for baskets anchored using 4 pins/basket and 10 pins/basket. Bar chart shows that 95.5 percent of dowels placed in 10-pin baskets were misaligned by 10 or fewer mm, compared to 68.6 percent of those placed in 4-pin baskets. Misalignments of 11-15 mm were found in 4.2 percent of the 10-pin placements, compared to 13.5 percent of the 4-pin placements. Misalignments of 16-20 mm were found in 0.2 percent of the 10-pin placements compared to 6.3 percent of the 4-pin placements. No misalignments greater than 20 mm were found in the 10-pin placements; however, misalignments of 21-25 mm and greater than 25 mm were found in 3.4 percent and 8.1 percent, respectively, of the 4-pin placements, indicating a high risk of distress.

Both the magnitude of misalignment and the number of dowel bars in each range of misalignment affect the functioning of a joint. Also, if a joint is locked due to one or more severely misaligned bars, the negative effect of additional misaligned bars may be minimal. In an attempt to quantify the potential effect of dowel misalignment on pavement performance, an index referred to as the "Joint Score" was developed to reflect the risk of joint locking. The index is determined based on the magnitude of misalignment and the number of dowel bars in each range of misalignment. An index of 10 or greater may suggest a high risk of joint locking. A full description of the index is provided in Appendix B. In Figures 39 and 40 a joint-by-joint evaluation of MIT Scan-2 data from the Nevada field test sections is shown using the Joint Score concept.

Figure 39. Joint-by-joint evaluation results for the baskets anchored using 4 pins per basket, showing a significant number of potentially locked joints (Joint Score > 10).
Joint-by-joint evaluation results for the baskets anchored using 4 pins per basket, showing a significant number of potentially locked joints (Joint Score > 10). The bar chart shows that 27 of 100 joints exceeded a Joint Score of 10.
Figure 40. Joint-by-joint evaluation results for the baskets anchored using 10 pins per basket, showing outstanding dowel alignment.
Joint-by-joint evaluation results for the baskets anchored using 10 pins per basket, showing outstanding dowel alignment. The bar chart shows that all of the 85 joints had Joint Scores of less than 5.

Figure 39 shows a high percentage of joints (27 percent) with a Joint Score greater than 10 for the section that was anchored using 4 pins per basket. The dramatic improvement in dowel alignment obtained when more pins were used (10 per basket) to hold the baskets securely in place is clearly shown in Figure 40. On this project, anchoring the baskets using 10 pins per basket completely eliminated the dowel misalignment problem. MIT Scan-2 was instrumental in both identifying the problem and verifying the improvement achieved after modifying the basket anchoring practice.

North Carolina

The North Carolina DOT was interested in a demonstration of MIT Scan-2 to determine its suitability for use during construction to monitor dowel alignment. The demonstration was particularly timely, because the DOT was contemplating whether to allow the contractor to use a DBI on an upcoming construction project (US-64 Knightdale Bypass). In the past, the DOT had not allowed the use of DBI. However, with the availability of practical means of verifying dowel alignment, the DOT was open to considering the DBI option.

Similar to other demonstrations, a short slide presentation was made, followed by an open-air demonstration. Shortly after this demonstration, the DOT approved the use of DBI with the condition that the contractor document dowel positions. On that project, both the contractor and the State used MIT Scan- 2 to monitor dowel alignment.

For the contractor, MIT Scan-2 was also instrumental in refining paving operations. Problems with equipment adjustment were easily detected using the MIT Scan-2 results. For example, a consistently large misalignment at one particular bar position suggested that the DBI forks at that position needed adjustment. For construction using a DBI, concrete mixture proportions have a significant effect on dowel placement. The portland cement concrete (PCC) mixture must be stable enough to hold the dowel bars in place after the insertion, and the mixture must be workable enough to ensure that no voids are left behind as the dowel bars pass through the concrete to their final positions. The ability to rapidly monitor dowel placement results at consecutive joints was helpful in adjusting the mixture proportions (Figure 41). This project demonstrated the usefulness of the MIT Scan-2 for process control in real time.

Figure 41. The effects of mix optimization on dowel alignment on the US-64 Bypass project.
 The effects of mix optimization on dowel alignment on the US-64 Bypass project. The bar chart shows a reduction in the percentage of misaligned bars on each of two succeeding paving dates.
Pennsylvania

MIT Scan-2 was demonstrated at a project site in Pennsylvania where cracking had developed on an I-81 reconstruction project. Because the cracking initiated at the ends of the outermost dowel, dowel misalignment was one of the suspect causes. Fifty joints were scanned using MIT Scan-2 in about an hour. The results showed that while the dowel alignment on this project is not exemplary, dowel misalignment is not likely to have been a contributing factor on the observed cracking.

The results are shown in Figures 42 and 43. As shown in Figure 42, this project contains what would appear to be a significant percentage of misaligned bars (misalignment > 20 mm [0.78 in.]). The Joint Scores for this project show that several joints on this project may be locked, but field experience shows that occasional locked joints can be tolerated (Yu 2005). Also, the joint locations on this project where cracking occurred did not have a high Joint Score.

Figure 42. Distribution of dowel misalignment in the I-81 project in Pennsylvania (maximum of either horizontal or vertical misalignment).
Distribution of dowel misalignment in the I-81 project in Pennsylvania (maximum of either horizontal or vertical misalignment) over five ranges: 10 mm or fewer, 62 percent; 11-15 mm, 24.1 percent; 16-20 mm, 8 percent; 21-25 mm, 3.8 percent; and more than 25 mm, 2.1 percent.
Figure 43. Joint Scores for the I-81 project in Pennsylvania.
Joint Scores for the I-81 project in Pennsylvania. The bar chart shows five bars with Joint Scores greater than 10.
<< PreviousContentsNext >>
 
Updated: 04/07/2011
 

FHWA
United States Department of Transportation - Federal Highway Administration